Working with Broken Phase-3A EEPROMs

Probe Labeling

4/30/2018 and later:

Prior to 4/30/2018:

Note: To program a broken probe we must know the full identifier printed on the probe and the option number. The serial number is essentially arbitrary. It is useful for communicating problems to Imec, and for your experiment metadata. Using "the wrong" 9 digits will not affect the probe or experiment outcomes.

EEPROM Chip

Probes have a tiny EEPROM chip (circled in red) containing the probe serial number, option, and ADC/gain calibration data.

EEPROM

EEPROM

If the chip detaches or breaks then SpikeGLX displays the following probe info when you click the 'Detect' button in the Devices tab:

  1. Note that true serial numbers never start with '1' and are only 9 digits long, whereas the bogus serial numbers usually start with '1' and are often longer.

  2. Although there are real option 4 probes, a broken EEPROM always presents as option 4.

The probe can still be used through a manual override feature. First ask the manufacturer for the calibration file set matching the probe's printed (nominally 11-digit) identifier. Then follow the instructions below to command SpikeGLX to use the files instead of the EEPROM chip.

Note: You will have to revisit the 'Force' dialog for this probe each time you quit and relaunch SpikeGLX.

Probe Data Folder

A probe's data folder looks like this:

    15039707141\
        Comparator calibration.csv
        Gain correction.csv
        Offset calibration.csv
        Slope calibration.csv

Install a Data Folder

If you need to access a broken probe, get its data folder and drag it into subfolder SpikeGLX/ImecProbeData.

If you don't see the ImecProbeData subfolder, create it yourself with exactly that case and spelling.

Using the "Force" Features

Click Detect on the Devices tab when you want to:

Click Same As Last Time to keep current flag settings for the next run.

Click Force... on the IM Setup tab to access data override features:

Code Entry in the Force dialog

Label: Enter the (nominally 11-digit) code printed on the probe, or click Explore ImecProbeData to copy it from the probe's folder so you can paste it into the box.

Serial #: Enter the 9-digit serial number directly, or click Take 9 From Label to create a 9-digit number based on the label.

Option: Remember to select the appropriate option number for this probe. This is not derived from the label as of 4/30/2018.

Skip ADC Calibration

This is not recommended, but available in cases where the probe data files are corrupt or you just don't have them and still want to use your broken probe.

fin