FIX: Imec Module Doesn't Work In My Chassis
You've powered everything off as instructed, installed your imec basestation module into a chassis slot and powered everything on again. You get a red module status light; the module isn't recognized at all...
It could be a 1-lane (X1) slot. In February 2022 we've encountered the fact that not all PXIe chassis slots are the same with respect to bandwidth. They come in X1, X2, X4, X8, X16, X32. X1 is sufficient for up to 250 MB/s, which is completely adequate for an imec module. However, imec originally designed the module firmware for X4 and higher slots.
Modules have two copies of the firmware. The normal/working copy is loaded, unless you power-up with the ISP button depressed, in which case the boot image is loaded. Note the boot image is a backup copy in case the working copy is corrupted.
In roughly Q2 2022, imec began shipping modules with an X1-compatible working image, but not an X1-compatible golden image. By Q2 2023, both images were X1-compatible.
What if you have an older module and a newer NI chassis, for example, a 1083 with five (X1) slots, a 1088 with five (X1) and three (X4) slots, or a Keysight M9005A that has only X1 slots? There's a firmware solution to fix the issue, but note the following:
-
You have to upgrade to a phase30 release; get the latest.
-
To update the module firmware for X1 compatibility, you first have to install it into an X4 or higher slot so you can talk to it...
-
Check if your chassis design includes at least some X4+ slots. It can be truly difficult to find that out. NI doesn't mention anything about that in most chassis 'Datasheets' or 'Specifications.' Rather, it's in the 'User Guide' for some models.
-
If you don't have any X4+ slots, maybe a neighbor lab can let you reprogram a module in their chassis.
The NI 1088 is a popular NI choice; the X1 slots are {2,3,5,7,9}, the X4 slots are {4,6,8}.
PXI Module Firmware
A module can only run one flavor of SpikeGLX at a time because it can only contain firmware of one flavor at a time. To run SpikeGLX-phase3B2 the hardware module must contain Neuropixels 1.0 (phase3B2) firmware. To run SpikeGLX-phase20 the module must contain Neuropixels 2.0 firmware. Likewise for the newest SpikeGLX-phase30.
You can always change the firmware, either to get the latest bug fixes, or to switch SpikeGLX flavors. Download the desired package (below) and unzip it to your hard drive. Note that each package contains two files, one for the FPGA on the basestation (BS) and one for the FPGA on the basestation connect card (BSC). The filenames contain either 'BS' or 'BSC' so you can tell them apart.
Run SpikeGLX and select menu item Tools/Update imec Firmware
. Click
the ? button in the dialog titlebar to get further help.
Regardless of the currently installed firmware, we recommend that you do ANY update using the latest SpikeGLX Phase30 application because that updater works reliably. Yes, even if for example, you are currently at Phase20 firmware and want to drop back to Phase3B2, we still advise you to run Phase30 software, select
Tools/Update imec Firmware
, then power everything down. After rebooting, run the SpikeGLX application that pairs with the newly loaded firmware.
fin